CSIR Central

A self assembled monolayer based microfluidic sensor for urea detection

IR@NPL: CSIR-National Physical Laboratory, New Delhi

View Archive Info
 
 
Field Value
 
Title A self assembled monolayer based microfluidic sensor for urea detection
 
Creator Srivastava, Saurabh
Solanki, Pratima R.
Kaushik, Ajeet
Ali, Md. Azahar
Srivastava, Anchal
Malhotra, B. D.
 
Subject Chemistry
Materials Science
Applied Physics/Condensed Matter
Nanoscience/ Nanotechnology
Physics
 
Description Urease (Urs) and glutamate dehydrogenase (GLDH) have been covalently co-immobilized onto a self-assembled monolayer (SAM) comprising of 10-carboxy-1-decanthiol (CDT) via EDC–NHS chemistry deposited onto one of the two patterned gold (Au) electrodes for estimation of urea using poly(dimethylsiloxane) based microfluidic channels (2 cm × 200 μm × 200 μm). The CDT/Au and Urs-GLDH/CDT/Au electrodes have been characterized using Fourier transform infrared (FTIR) spectroscopy, contact angle (CA), atomic force microscopy (AFM) and electrochemical cyclic voltammetry (CV) techniques. The electrochemical response measurement of a Urs-GLDH/CDT/Au bioelectrode obtained as a function of urea concentration using CV yield linearity as 10 to 100 mg dl−1, detection limit as 9 mg dl−1 and high sensitivity as 7.5 μA mM−1 cm−2.
 
Publisher Royal Society of Chemistry
 
Date 2011-04-15
 
Type Article
PeerReviewed
 
Format application/pdf
 
Identifier http://npl.csircentral.net/778/1/7.pdf
Srivastava, Saurabh and Solanki, Pratima R. and Kaushik, Ajeet and Ali, Md. Azahar and Srivastava, Anchal and Malhotra, B. D. (2011) A self assembled monolayer based microfluidic sensor for urea detection. Nanoscale , 3 (7). pp. 2971-2977. ISSN 2040-3372
 
Relation http://dx.doi.org/10.1039/C1NR10240K
http://npl.csircentral.net/778/