Studies on the stabilities of various types of industrial pressure measuring devices

IR@NPL: CSIR-National Physical Laboratory, New Delhi

View Archive Info
 
 
Field Value
 
Title Studies on the stabilities of various types of industrial pressure measuring devices
 
Creator Yadav, Sanjay
Prakash, Om
Gupta, V. K.
Bandyopadhyay, A. K.
 
Subject Engineering
 
Description Present paper describes studies carriled out on an industrial dead weight tester (pressure range, up to 330 MPa; calibration; 3 times: period, 6 years), one pressure dial gauge (pressure range, up to 70 MPa; calibration, 3 time period, 3 years), and one digital pressure calibrator (pressure range, up to 200 MPa; calibration, 4 times; period, 6 years). A small change was observed in the short-term stability but the change was prominent in case of long-term stability in all the instruments. Stability of dead weight tester and pressure dial gauge was well within their claimed manfacturer Uncertain ties while it was found to be beyond claimed uncertainty in case of digital pressure calibrator.
 
Publisher NISCAIR
 
Date 2006-09
 
Type Article
PeerReviewed
 
Format application/pdf
 
Identifier http://npl.csircentral.net/2561/1/154.pdf
Yadav, Sanjay and Prakash, Om and Gupta, V. K. and Bandyopadhyay, A. K. (2006) Studies on the stabilities of various types of industrial pressure measuring devices. Journal of Scientific and Industrial Research, 65 (9). 721-724 . ISSN 0022-4456
 
Relation http://npl.csircentral.net/2561/