Molecularly imprinted polyaniline-polyvinyl sulphonic acid composite based sensor for para-nitrophenol detection
IR@NPL: CSIR-National Physical Laboratory, New Delhi
View Archive InfoField | Value | |
Title |
Molecularly imprinted polyaniline-polyvinyl sulphonic acid composite based sensor for para-nitrophenol detection
|
|
Creator |
Roy, Abhijit Chandra
Nisha, V. S. Dhand, Chetna Ali, Md Azahar Malhotra, B. D. |
|
Subject |
Analytical Chemistry
|
|
Description |
We report results of the studies relating to the fabrication and characterization of a conducting polymer based molecularly imprinted para-nitrophenol (PNP) sensor. A water pollutant, para-nitrophenol is electrochemically imprinted with polyvinyl sulphonic acid (PVSA) doped polyaniline onto indium tin oxide (ITO) glass substrate. This PNP imprinted electrode (PNPI-PANI-PVSA/ITO) prepared via chronopotentiometric polymerization and over-oxidation is characterized by Fourier transform infra-red spectroscopy (FT-IR), UV-visible (UV-vis) spectroscopy, contact angle (CA), scanning electron microscopy (SEM), cyclic voltammetry (CV) and differential pulse voltammetry (DPV) studies. The response studies of PNPI-PANI-PVSA/ITO electrode carried out using DPV reveal a lower detection limit of 1 x 10(-3) mM, improved sensitivity as 1.5 x 10(-3) A mM(-1) and stability of 45 days. The PNPI-PANI-PVSA/ITO electrode shows good precision with relative standard deviation of 2.1% and good reproducibility with standard deviation of 3.78%.
|
|
Publisher |
Elsevier
|
|
Date |
2013-03-13
|
|
Type |
Article
PeerReviewed |
|
Format |
application/pdf
|
|
Identifier |
http://npl.csircentral.net/3120/1/Molecularly%20imprinted%20polyaniline.pdf
Roy, Abhijit Chandra and Nisha, V. S. and Dhand, Chetna and Ali, Md Azahar and Malhotra, B. D. (2013) Molecularly imprinted polyaniline-polyvinyl sulphonic acid composite based sensor for para-nitrophenol detection. Analytica Chimica Acta , 777. pp. 63-71. ISSN 0003-2670 |
|
Relation |
http://npl.csircentral.net/3120/
|
|